Stack Emission Methods: EPA, EN and HJ Equivalence

Which method applies to your report, and which instrument executes it. This is the reference we use internally, written out so the answer does not depend on a sales conversation.

This page describes method scope and standard numbering. It is not a claim of regulatory equivalence. Requirements for probe construction, flow verification, leak checks, absorption solutions and calculation conventions differ between frameworks. Kesen supplies the instrument configuration and verification records appropriate to the framework you specify; the compliance report is prepared by your own accredited test body. Verify every reference below before publishing.

Why the method number matters more than the instrument

An instrument does not “comply with EPA Method 1-5”. The test complies with Method 1-5. Regulatory bodies assess the sampling procedure, the verification records and the calculation — not the instrument’s brand. What a manufacturer can legitimately do is build an instrument whose probe geometry, flow control, filtration and data capture meet the method’s requirements, and supply the verification records that demonstrate it.

This distinction matters commercially. A supplier who claims an instrument “meets EPA 1-5” without naming what it does and does not satisfy is asking you to take their word for the part that matters. A supplier who says “this is the principle, these are the standards the instrument is built to, and this is what your test body still has to do” is giving you what you need to specify correctly.

The equivalence table

Each row lists a parameter, the method framework in each jurisdiction, what the method actually determines, and the Kesen instruments that execute it. Every model code links to its product page.

Parameter China United States Europe (EN) Germany What the method determines · instruments
Particulate matter, isokinetic (gravimetric)
VDI 4206 Blatt 2
GB/T 16157-1996 · HJ/T 48-1999 US EPA Method 1–5 EN 13264-1 VDI 4206 Blatt 2 Mass concentration, equivalent concentration and total emission of particulate matter in stationary source emissions, by isokinetic sampling, filtering and gravimetric weighing.
3012H-D · 3012H · 3012H-C · 3012H-Y · 3012B
Low-concentration particulate matter
VDI 4206 Blatt 2
DB37/T 2537-2014 (and 7 other provincial methods) US EPA Method 1–5, modified probe EN 13264-1 low-level variant VDI 4206 Blatt 2 Gravimetric determination where concentration falls below 50 mg/m³. Our 3012H-D states a lower testing limit of 1 mg/m³ at a 1 m³ standard-condition sample volume.
3012H-D · 3012H-C · 1089K
Stack gas sampling conditions
VDI 4206 Blatt 1
HJ/T 47-1999 US EPA Method 3 / 3A EN 13264-1 (velocities) VDI 4206 Blatt 1 Flow rate, velocity, pressure and temperature at the sampling point, establishing the isokinetic reference.
3012H-D · 3012H · 3060-A · 1062E
SO₂
VDI 4206 Blatt 1
HJ 1131-2020 · HJ 57-2017 · HJ 1045-2019 US EPA Method 6 / 6A EN 15259-3 VDI 4206 Blatt 1 Sulphur dioxide by constant-potential electrolysis or iodometric titration, referenced to measured oxygen.
3023 · 3023Y · 3021 · 2086
NOₓ
VDI 4206 Blatt 1
HJ 1132-2020 US EPA Method 7 / 7A EN 15259-3 VDI 4206 Blatt 1 Nitrogen oxides by constant-potential electrolysis, reported as NO, NO₂ or NOₓ against a stated oxygen reference.
3023 · 3023Y · 2086
CO
VDI 4206 Blatt 1
HJ/T 44-1999 US EPA Method 10A EN 15259-3 VDI 4206 Blatt 1 Carbon monoxide by constant-potential electrolysis, measured with oxygen for reference-condition correction.
3021 · 2086
O₂ reference concentration
VDI 4206 Blatt 1
HJ/T 42-1999 US EPA Method 3A EN 15259-3 VDI 4206 Blatt 1 Oxygen as the reference parameter converting measured concentrations to a defined reference condition. Every corrected result depends on it.
3012H-D · 3012H · 3023 · 3023Y · 2086
Flue gas velocity and moisture
VDI 4206 Blatt 1
HJ/T 47-1999 US EPA Method 2 / 2F EN 15259-2 VDI 4206 Blatt 1 Volumetric flow from pitot and velocity measurement. Moisture determines whether a result is reported dry or wet, and the two can differ enough to fail a limit.
3060-A · 1062C · 1062E · 3012H-D
Mercury in stack gas
VDI 4206 Blatt 1
HJ 57-2017 · HJ 543-2009 US EPA Method 30A / 30B EN 13264-1 with CEN procedure VDI 4206 Blatt 1 Mercury by cold atomic absorption spectrophotometry, sampled on a dual-channel train so mercury and conventional gas species can be captured on the same run.
3072 · 3023Y · 2086
Dioxins and furans in stack gas
—
GB 16445-2014 · HJ/T 295-2009 US EPA Method 23 EN 1948-2 — PCDD/F and PCDF by high-resolution mass spectrometry after combined sampling. Sampling-limited, not analysis-limited: the result is set by the volume you can draw cleanly.
3030B
Dioxins in ambient air
—
HJ/T 194-2017 US EPA Method 1A EN 1948-2 — Ambient dioxin sampling on a high-volume filter train.
Volatile organic compounds in exhaust gas
—
HJ 38-2017 · HJ/T 732-2014 US EPA Method 002/003 class EN 15259-6 — Canister, bag or variable-temperature condensation sampling, followed by GC/MS analysis.
3036 · 3038C · 2083
VOC in ambient air
—
HJ 644-2013 · HJ 645-2013 — EN ISO 2022 class — Sorbent tube thermal desorption, or activated charcoal with carbon disulphide desorption, followed by GC/MS.
2061 · 2033B
Fugitive emissions and leaks
—
GB 37822-2019 · HJ 733-2019 — — — Photoionisation detection for LDAR programmes. A PID is the working reference for leak screening; it is not a substitute for a regulatory total-hydrocarbon method.
3033 · 3233 · 2028
Total hydrocarbon and methane
—
HJ 1012-2018 · HJ 1331-2023 — — — Total hydrocarbon, methane and non-methane total hydrocarbon by FID, by catalytic oxidation, GC separation or GC column.
3035 · 3033
Continuous particulate monitoring (CEMS)
17. BImSchV + QAL2T
HJ/T 75-2017 · HJ 75-2007 US EPA Method 101 / 101A, 40 CFR Part 75 EN 15259-2 17. BImSchV + QAL2T Unattended beta-ray or equivalent continuous measurement. The data availability percentage is a compliance requirement in its own right, not just a quality metric.
1089K
Gaseous pollutants (online systems)
39. BImSchV + VDI 4202
HJ 1100-2020 · HJ 653-2013 US EPA 40 CFR 50 Directive 2008/50/EC 39. BImSchV + VDI 4202 Automated gaseous pollutant measurement systems for ambient networks.
2092 · 2028
Data transmission for network systems
VDI 4201 Blatt 1–4
HJ/T 212-2005 · HJ/T 212-2017 EPA AQS — VDI 4201 Blatt 1–4 The transmission protocol municipal and regulatory networks are required to speak. A monitor that cannot speak it needs a separate data logger and software.
2092 · 2193
Environmental noise
ISO 61672 + DIN 45645
GB 3096-2008 US EPA 42 CFR 49 ISO 3744 / ISO 9612 ISO 61672 + DIN 45645 Environmental noise by a Class 1 sound level meter, with the required statistical descriptors including day/night levels.
2193
Odor
—
GB 14554-93 — EN 13725 (triangle odour bag) — Odor concentration by dynamic olfactometry. The value is determined by a trained sensory panel, not by an instrument reading alone.
8032 · 2022
COD / chemical oxygen demand
—
HJ 828-2017 US EPA 40 CFR 261.4 / 405.0 ISO 11909-2 class — Dichromate oxidation with closed reflux. The regulatory method; a permanganate index is a different measurement, not an alternative.
5034
Permanganate index
—
GB 11892-89 · GB/T 5750.7-2023 — — — A weaker oxidation suited to surface water and drinking water monitoring where high COD is not expected.
5032
Flow calibration verification
—
JJG 586-2006 · JJG 640-2016 · JJG 875-2019 — — — Traceable verification of the flow path a compliance measurement depends on. An untraceable calibration breaks the chain retrospectively.
7020A · 7030 · 7050B · 8040

Which framework applies to your report?

  • Operating across multiple jurisdictions? Specify per jurisdiction and document the mapping internally. A configuration that satisfies one framework is frequently not optimal for another, and the difference is usually range rather than principle.
  • Unsure which implementing method applies? Ask your national environmental authority, not the instrument supplier. The authority publishes the applicable method; a supplier can only tell you what their instrument is built to.
  • Third-party testing for a client? The client or their regulator specifies the framework. Confirm it in writing before quoting — a change of framework after the test invalidates the work.

Configuration decisions that change the result

Low-concentration particulate

Below roughly 50 mg/m³ the gravimetric result approaches the method blank. A standard configuration can produce a number that looks valid and is not defensible. The low-concentration multi-function probe and a higher-flow pump exist for exactly this case. Our 3012H-D states a lower testing limit of 1 mg/m³ at a 1 m³ standard-condition sample volume. This is a configuration decision at order, not an upgrade after delivery.

Moisture basis

Stack gas results are reported dry or wet depending on the method, and the two can differ by enough to fail a limit. Moisture is not a housekeeping parameter — it enters the calculation. Measure it in the same run rather than carrying a value over from a previous test.

Probe selection

Probe diameter, material and temperature rating must match the duct geometry and the flue gas temperature. A leak in the probe or a degraded filter holder biases the entire result, and the bias may not be detectable in the subsequent calculation.

Which gases, and in what range

The low-range and high-range SO² configurations of the same analyser are different instruments in practice. State the expected concentration range in your enquiry so the configuration is confirmed in the quotation rather than discovered on site.

Download the table

standards-equivalence-table.csv — the full table as a spreadsheet, one row per parameter, with the China, US, EU and Germany references and the Kesen models that execute each method.

Machine-readable and free to redistribute with attribution. If you are compiling a method list for a tender or a client submission, this is the version to attach.

Frequently asked questions

An instrument does not “comply with EPA Method 1-5”. The test does.

A regulatory body assesses the sampling procedure, the verification records and the calculation — not the instrument's brand. What a manufacturer can legitimately do is build an instrument whose probe geometry, flow control, filtration and data capture meet the method's requirements, and supply the verification records. Kesen states plainly which frameworks an instrument is built to, and what your accredited test body still has to do.

Can I report an EPA Method 1-5 result using a Chinese-standard sampler?

The physical principle, probe geometry and gravimetric determination are the same. What differs is procedural detail: flow-rate verification method, leak-check procedure, probe construction tolerances and calculation conventions. An instrument built to GB/T 16157-1996 and HJ/T 48-1999 addresses the same measurement. Confirm the specific differences your authority expects with your own accredited tester — before the test, not after.

Which framework should I specify?

The framework your regulator enforces — in practice the national implementing method of the jurisdiction where the report is filed, not the method you are most familiar with. If you operate across jurisdictions, specify per jurisdiction and document the mapping internally; one configuration frequently cannot satisfy two frameworks at once, and the difference is usually range rather than principle.

What is the difference between a standard and a low-concentration configuration?

Below roughly 50 mg/m³ of particulate, the gravimetric result approaches the method blank and a standard configuration can no longer produce a defensible number. The low-concentration configuration uses a higher-flow pump and a multi-function probe to increase the sampled mass. Our 3012H-D states a lower testing limit of 1 mg/m³ at a 1 m³ standard-condition sample volume. This is a configuration decision at order, not an upgrade after delivery.

Should I report dry or wet gas concentration?

Whichever the method requires — and the two can differ by enough to fail a limit. Moisture is not a housekeeping parameter; it enters the calculation. Measure it in the same run using the 1062C or 1062E rather than carrying a value over from a previous test.

Do you supply the verification records my accredited body needs?

Each instrument is calibrated against traceable reference standards before dispatch and ships with a calibration record showing the measured values, the reference standard used and the acceptance tolerance. Verification records are provided per the applicable JJG regulation. We supply the instrument configuration and documentation; the compliance report is prepared by your own accredited test body.

What is the real difference between an NDIR and an NDUV analyser?

The measurement principle and the application. NDIR uses short-wave infrared absorption and suits CO, CO₂, SO₂, CH₄, NOₓ and NH₃. NDUV uses ultraviolet absorption and suits O₃, NO, NO₂ and SO₂. Both are non-dispersive, which removes moving parts. The choice comes down to which gases you need and the concentration range, not to which technology is better.

Are your instruments accurate to the figure you publish?

We publish measurement conditions with every figure, because an accuracy figure without its range, gas, ambient temperature, compensation mode and response time is not a usable specification. If a competitor's datasheet omits those conditions, ask both of us for the same test under the same conditions and compare.

Can one configuration cover several method frameworks?

Frequently the instrument can, which is why the multi-gas analysers are built as configurable platforms. But a single configuration usually cannot satisfy every framework simultaneously — a low-range and a high-range SO₂ measurement are different configurations of the same analyser. Ask for the configuration to be confirmed in the quotation against the frameworks you actually work under.